Simple and universal HPLC-UV method to determine cimetidine, ranitidine, famotidine and nizatidine in urine: Application to the analysis of ranitidine and its metabolites in human volunteers
- ,
- Rajesh Patel,
- Abdul W. Basit(corresponding author)
- University College London,
- GSK R and D Ltd.
Research Output:
Contribution to journal
Article
Research
Peer-reviewAbstract
<p>A validated, simple and universal HPLC-UV method for the determination of cimetidine, famotidine, nizatidine and ranitidine in human urine is presented. This is the first single HPLC method reported for the analysis of all four H<sub>2</sub> antagonists in human biological samples. This method was also utilized for the analysis of ranitidine and its metabolites in human urine. All calibration curves showed good linear regression (r<sup>2</sup> > 0.9960) within test ranges. The method showed good precision and accuracy with overall intra- and inter-day variations of 0.2-13.6% and 0.2-12.1%, respectively. Separation of ranitidine and its metabolites using this assay provided significantly improved resolution, precision and accuracy compared to previously reported methods. The assay was successfully applied to a human volunteer study using ranitidine as the model compound.</p>
Research Output Information
Output type
Research Output:
Contribution to journal
Article
Research
Peer-reviewOriginal language
EnglishPages from-to (Number of pages)
Pages 235-240 (6 pages)Journal (Volume, Issue Number)
Journal of Chromatography B: Analytical Technologies in the Biomedical and Life Sciences (Volume 860, Issue 2)Publication milestones
- Published - 15/12/2007
Publication status
Published - 15/12/2007
ISSN
1570-0232Publication IDs
- Scopus: 36549032068
- PubMed: 18006395