Abstract

Real-time PCR was employed to detect a region of the P1 cytadhesin gene of Mycoplasma pneumoniae in clinical samples. An internal processing control was included that could be co-amplified simultaneously in the same reaction tube. The assay could reproducibly detect 1 × 103 M. pneumoniae organisms ml-1 in clinical samples. There was no amplification of DNA or signal production from 15 other species of human mycoplasmas and 19 other bacterial species. Using a panel of 175 respiratory samples taken from patients with pneumonia of proven aetiology, the sensitivity was found to be 60% and the specificity of the assay 96.7% when compared with serology. This assay is suitable for same-day diagnosis of M. pneumoniae infection and batch processing of respiratory samples for clinical screening.

Research Output Information

Output type

Research Output:
Contribution to journal
Article
Research
Peer-review

Original language

English

Pages from-to (Number of pages)

Pages 149-155 (7 pages)

Journal (Volume, Issue Number)

Journal of Medical Microbiology (Volume 55, Issue 2)

Publication milestones

  • Published - 02/2006

Publication status

Published - 02/2006

ISSN

0022-2615

Publication IDs

  • Scopus: 32044440317
  • PubMed: 16434706