Skip to search boxSkip to navigationSkip to main content

Investigating the antimicrobial efficacy of two aerosolized hydrogen peroxide fumigation cycles for biological safety cabinet decontamination

*Corresponding author for this work
Research Output:
Contribution to journal
Article
Research
Peer-review

Open access

Abstract

Aims: Fumigation of biological safety cabinets (BSCs) is essential for their safe use, maintenance, and quality assurance of work performed. Hydrogen peroxide (HP) is frequently used as an alternative fumigant to formaldehyde, applied either at high concentrations (>30%) in vapour phase, or low concentrations (5%–12%) as an aerosol (aHP). There is, however, limited evidence on the efficacy of low-concentration aerosolized HP (aHP) fumigation to inactivate microorganisms within BSCs. This study evaluates the antimicrobial efficacy of aHP within a BSC III, comparing fumigation cycle parameters against a range of microorganisms. Methods and results: Biological indicators (BIs) were prepared by inoculating stainless steel coupons with either Geobacillus stearothermophilus spores, methicillin-resistant Staphylococcus aureus (MRSA), or Mycobacterium fortuitum. BIs were placed within a BSC III and fumigated with one or six diffusion cycles of aHP, with one diffusion cycle recommended for routine decontamination and six diffusion cycles advised for pre-maintenance decontamination, providing increased aHP quantity and exposure time. Fumigation with one aHP diffusion achieved a > 6 log10 reduction in G. stearothermophilus spore CFU, with 2.24 and 0.46 log10 average reductions for MRSA and M. fortuitum, respectively. Six-diffusion fumigation achieved significantly reduced CFU counts for all microorganisms (average log10 reductions of 7.05, 5.25, and 5.26 for G. stearothermophilus, MRSA, and M. fortuitum, respectively). Incomplete BI inactivation was, however, observed for all microorganisms. Conclusions: Increased aHP diffusion cycles enhanced microbial reductions; however, complete elimination of microorganisms was not observed. BIs, which represent likely contaminated microorganisms should be used when validating fumigation effectiveness.

Research Output Information

Output type

Research Output:
Contribution to journal
Article
Research
Peer-review

Original language

English

Article number

lxag154

Journal (Volume, Issue Number)

Journal of Applied Microbiology (Volume 137, Issue 7)

Publication milestones

  • Published - 07/2026

Publication status

Published - 07/2026

ISSN

1364-5072

Publication IDs

  • Scopus: 105044412904

Funding Details

This work was supported by the UKHSA Grant-in-Aid funding. The views expressed in this manuscript are those of the authors, not those of UKHSA or other funding sources. Crown Copyright. UK Health Security Agency, 2026.
FunderFunding number
UKHSA
Grant-in-Aid funding