Abstract
A procedure is described for re-assessing the dose from the Vinten extremity dosemeter using phototransferred thermoluminescence. The technique produces a linear response up to 50 Sv. The re-assessment efficiency is approximately 10% for the first re-assessment and 3-5% for the second re-assessment. The detection threshold values (at the 95% confidence level) are 3 mSv and 9 mSv respectively.
| Original language | English |
|---|---|
| Pages (from-to) | 119-120 |
| Number of pages | 2 |
| Journal | Radiation Protection Dosimetry |
| Volume | 27 |
| Issue number | 2 |
| Publication status | Published - 1989 |