Re-assessment of dose from the Vinten extremity dosemeter

J. B. O'Hagan, A. J. Pearson, J. C. Dutt

Research output: Contribution to journalArticlepeer-review

1 Citation (Scopus)

Abstract

A procedure is described for re-assessing the dose from the Vinten extremity dosemeter using phototransferred thermoluminescence. The technique produces a linear response up to 50 Sv. The re-assessment efficiency is approximately 10% for the first re-assessment and 3-5% for the second re-assessment. The detection threshold values (at the 95% confidence level) are 3 mSv and 9 mSv respectively.

Original languageEnglish
Pages (from-to)119-120
Number of pages2
JournalRadiation Protection Dosimetry
Volume27
Issue number2
Publication statusPublished - 1989

Fingerprint

Dive into the research topics of 'Re-assessment of dose from the Vinten extremity dosemeter'. Together they form a unique fingerprint.

Cite this