T1 - Newborn screening for sickling and other haemoglobin disorders using tandem mass spectrometry
T2 - A pilot study of methodology in laboratories in England
AU - Daniel, Yvonne A.
AU - Henthorn, Joan
N1 - Publisher Copyright:
© The Author(s) 2016.
Copyright:
Copyright 2021 Elsevier B.V., All rights reserved.
PY - 2016/12
Y1 - 2016/12
N2 - Objective: To determine (i) if electrospray mass spectrometry-mass spectrometry with the SpOtOn Diagnostics Ltd reagent kit for sickle cell screening could be integrated into the English newborn screening programme, under routine screening conditions, and provide mass spectrometry-mass spectrometry results which match existing methods, and (ii) if common action values could be set for all manufacturers in the study, for all assessed haemoglobins, to indicate which samples require further investigation. Methods: Anonymised residual blood spots were analysed using the SpOtOn reagent kit as per manufacturer’s instructions, in parallel with existing techniques at four laboratories. Mass spectrometry-mass spectrometry instrumentation at Laboratories A and B was AB Sciex (Warrington, UK) AP4000, and at Laboratories C and D,Waters Micromass (Manchester, UK), Xevo TQMS and Premier, respectively. Results: There were 23,898 results accepted from the four laboratories. Excellent specificity at 100% sensitivity was observed for haemoglobin S, haemoglobin C, haemoglobin E and haemoglobin OArab. A common action value was not possible for Hb C, but action values were set by manufacturer. The two haemoglobin DPunjab cases at Laboratory D were not detected using the common action value. Conversely, false-positive results with haemoglobin DPunjab were a problem at the remaining three laboratories. Conclusions: This multicentre study demonstrates that it is possible to implement mass spectrometry-mass spectrometry into an established screening programme while maintaining consistency with existing methods for haemoglobinopathy screening. However, one of the instruments investigated cannot be recommended for use with this application.
AB - Objective: To determine (i) if electrospray mass spectrometry-mass spectrometry with the SpOtOn Diagnostics Ltd reagent kit for sickle cell screening could be integrated into the English newborn screening programme, under routine screening conditions, and provide mass spectrometry-mass spectrometry results which match existing methods, and (ii) if common action values could be set for all manufacturers in the study, for all assessed haemoglobins, to indicate which samples require further investigation. Methods: Anonymised residual blood spots were analysed using the SpOtOn reagent kit as per manufacturer’s instructions, in parallel with existing techniques at four laboratories. Mass spectrometry-mass spectrometry instrumentation at Laboratories A and B was AB Sciex (Warrington, UK) AP4000, and at Laboratories C and D,Waters Micromass (Manchester, UK), Xevo TQMS and Premier, respectively. Results: There were 23,898 results accepted from the four laboratories. Excellent specificity at 100% sensitivity was observed for haemoglobin S, haemoglobin C, haemoglobin E and haemoglobin OArab. A common action value was not possible for Hb C, but action values were set by manufacturer. The two haemoglobin DPunjab cases at Laboratory D were not detected using the common action value. Conversely, false-positive results with haemoglobin DPunjab were a problem at the remaining three laboratories. Conclusions: This multicentre study demonstrates that it is possible to implement mass spectrometry-mass spectrometry into an established screening programme while maintaining consistency with existing methods for haemoglobinopathy screening. However, one of the instruments investigated cannot be recommended for use with this application.
KW - Haemoglobinopathy
KW - Newborn screening
KW - Sickle cell disease
KW - Tandem mass spectrometry
UR - http://www.scopus.com/inward/record.url?scp=85011656488&partnerID=8YFLogxK
U2 - 10.1177/0969141316631008
DO - 10.1177/0969141316631008
M3 - Article
C2 - 26965059
AN - SCOPUS:85011656488
VL - 23
SP - 175
EP - 178
JO - Journal of Medical Screening
JF - Journal of Medical Screening
SN - 0969-1413
IS - 4
ER -