Abstract
Performance characterisations were carried out before and after a modification to the optics of the Condor Instruments’ ActTrust light and activity data loggers to improve the spectral performance for measuring melanopic-weighted irradiance in non-visual studies. The results confirm the intended improvement, so that the device provides the best-known single-sensor match to the melanopic response. In addition, the device includes a separate sensor which remained well-matched for illuminance logging.