Abstract
The performance of the germanium (Ge) detector body monitor that has been recently installed at NRPB Chilton is evaluated and compared with the performance of the existing NRPB scintillation detector body monitor. Calibration of a p-type Ge detector for the measurement of radionuclides distributed throughout the body which emit photons with energies >100 keV is described. An ongoing systematic calibration programme of n-type Ge detectors for the measurement of low energy (<200 keV) photon emitting radionuclides in the lung is outlined. Results for the detection sensitivity for 241Am in the right lung are presented.
| Original language | English |
|---|---|
| Pages (from-to) | 73-75 |
| Number of pages | 3 |
| Journal | Radiation Protection Dosimetry |
| Volume | 53 |
| Issue number | 1-4 |
| DOIs | |
| Publication status | Published - 1994 |
UN SDGs
This output contributes to the following UN Sustainable Development Goals (SDGs)
-
SDG 3 Good Health and Well-being
Fingerprint
Dive into the research topics of 'Evaluation of a high purity germanium detector body monitor'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver