TY - JOUR
T1 - Application of Monte Carlo calculations to calibration of anthropomorphic phantoms used for activity assessment of actinides in lungs
AU - Franck, Didier
AU - Borissov, N.
AU - de Carlan, L.
AU - Pierrat, N.
AU - Genicot, J. L.
AU - Etherington, George
N1 - Copyright:
Copyright 2018 Elsevier B.V., All rights reserved.
PY - 2003
Y1 - 2003
N2 - This paper reports on a new utility for development of computational phantoms for Monte Carlo calculations and data analysis for in vivo measurements of radionuclides deposited in tissues. The individual parameters of each worker can be acquired for an exact geometric representation of his or her anatomy, which is particularly important for low-energy gamma ray emitting sources such as thorium, uranium, plutonium and other actinides. The software discussed here enables automatic creation of an MCNP input data file based on computed tomography (CT) scanning data. The utility was first tested for low- and medium-energy actinide emitters on Livermore phantoms, the mannequins generally used for lung counting, in order to compare the results of simulation and measurement. From these results, the utility's ability to study uncertainties in in vivo calibration were investigated. Calculations and comparison with the experimental data are presented and discussed in this paper.
AB - This paper reports on a new utility for development of computational phantoms for Monte Carlo calculations and data analysis for in vivo measurements of radionuclides deposited in tissues. The individual parameters of each worker can be acquired for an exact geometric representation of his or her anatomy, which is particularly important for low-energy gamma ray emitting sources such as thorium, uranium, plutonium and other actinides. The software discussed here enables automatic creation of an MCNP input data file based on computed tomography (CT) scanning data. The utility was first tested for low- and medium-energy actinide emitters on Livermore phantoms, the mannequins generally used for lung counting, in order to compare the results of simulation and measurement. From these results, the utility's ability to study uncertainties in in vivo calibration were investigated. Calculations and comparison with the experimental data are presented and discussed in this paper.
UR - http://www.scopus.com/inward/record.url?scp=0141563799&partnerID=8YFLogxK
U2 - 10.1093/oxfordjournals.rpd.a006269
DO - 10.1093/oxfordjournals.rpd.a006269
M3 - Article
C2 - 14526997
AN - SCOPUS:0141563799
SN - 0144-8420
VL - 105
SP - 403
EP - 408
JO - Radiation Protection Dosimetry
JF - Radiation Protection Dosimetry
IS - 1-4
ER -