Abstract
Seven patients developed Pseudomonas aeruginosa urinary tract infections following urodynamic studies, over a two-month period. One patient developed septicaemia and meningitis and died following rupture of a berry aneurysm. Two others required hospital admission for intravenous antibiotic treatment. Pseudomonas aeruginosa was isolated from a pressure dome which covered the pressure transducer of the urodynamic system used for assessing bladder pressure. The device packaging carried the symbol designating this as a 'single use' product, but as an economy measure, a local decision had been taken to change the device monthly.
| Original language | English |
|---|---|
| Pages (from-to) | 60-63 |
| Number of pages | 4 |
| Journal | Journal of Hospital Infection |
| Volume | 47 |
| Issue number | 1 |
| DOIs | |
| Publication status | Published - 2001 |
| Externally published | Yes |
UN SDGs
This output contributes to the following UN Sustainable Development Goals (SDGs)
-
SDG 3 Good Health and Well-being
Keywords
- Pseudomonas aeruginosa
- Urinary tract infections
Fingerprint
Dive into the research topics of 'An outbreak of Pseudomonas aeruginosa infection associated with contaminated urodynamic equipment'. Together they form a unique fingerprint.Cite this