An outbreak of Pseudomonas aeruginosa infection associated with contaminated urodynamic equipment

  • D. P.J. Turner
  • , T. C. Boswell
  • , J. V. Lee
  • , R. C.B. Slack
  • , R. P. Burden

Research output: Contribution to journalArticlepeer-review

22 Citations (Scopus)

Abstract

Seven patients developed Pseudomonas aeruginosa urinary tract infections following urodynamic studies, over a two-month period. One patient developed septicaemia and meningitis and died following rupture of a berry aneurysm. Two others required hospital admission for intravenous antibiotic treatment. Pseudomonas aeruginosa was isolated from a pressure dome which covered the pressure transducer of the urodynamic system used for assessing bladder pressure. The device packaging carried the symbol designating this as a 'single use' product, but as an economy measure, a local decision had been taken to change the device monthly.

Original languageEnglish
Pages (from-to)60-63
Number of pages4
JournalJournal of Hospital Infection
Volume47
Issue number1
DOIs
Publication statusPublished - 2001
Externally publishedYes